Concept
TriOS RAMSES ACC-UV Hyperspectral UVA and UVB Radiometer
URI | http://vocab.nerc.ac.uk/collection/L22/current/TOOL0810/ | ||
---|---|---|---|
Within Vocab | SeaVoX Device Catalogue | ||
Alternative Labels | RAMSES ACC-UV radiometer | ||
Definition | This is a spectral radiometer measuring over the UV band range (280-500 nm) with a cosine-response to measure irradiance. It is fitted with a 256 channel silicon photodiode array with channels giving 2.2 nm/pixel and a spectral accuracy of 02 nm. Typical saturation (4ms integration time) is 20 W/m2/nm/sr (at 300 nm), 17 W/m2/nm/sr (at 360nm), 18W W/m2/nm/sr (at 500nm). Typical NEI (8ms integration time) is 0.85 microW/m2/nm/sr (at 300nm), 0.75 microW/m2/nm/sr (at 360nm), 0.80 microW/m2/nm/sr (at 500nm). Accuracy is quoted as being better than 6-10 per cent depending on spectral range. | ||
Date | 2014-12-08T15:22:19 | ||
Identifier | SDN:L22::TOOL0810 | ||
Note | accepted | ||
Has Current Version | 1 | ||
version | 1 | ||
inScheme | http://vocab.nerc.ac.uk/scheme/SDNDEV/current/ | ||
Broader | L05:122 |
radiometers | Mapping: 540945 |
L19:SDNKG17 |
sensor_model | Mapping: 1696639 | |
Related | B75:ORG00579 |
TriOS Mess- und Datentechnik GmbH | Mapping: 1448852 |
Alternate Formats
Other formats for this page:
RDF/XML Turtle JSON-LDAlternate Profiles
Other views of this page:
Alternate Profiles ?Different Media Types (HTML, text, RDF, JSON etc.) and different information model views, profiles, are available for this resource.
NVS html view ? Default NVS html view.