Concept
Valeport miniCTD DR CTD
URI | http://vocab.nerc.ac.uk/collection/L22/current/TOOL1437/ | ||
---|---|---|---|
Within Vocab | SeaVoX Device Catalogue | ||
Alternative Labels | Valeport miniCTD-DR | ||
Definition | A CTD that samples by time (continuous) or depth (profiling). It is designed for ROV applications. It uses a platinum resistance thermometer, strain gauge pressure transducer (from a selection of measurement ranges between 0-600 Bar) and pressure-balanced conductivity cell. The instrument has a direct reading interface only and no internal battery or logging capability. It is housed in titanium. Data may be sampled from 1 to 8Hz. It is rated to 6000 m, although the measurement range is limited by the range of the pressure transducer used. Temperature accuracy is +/-0.01 degC. Conductivity accuracy is +/- 0.01 mS/cm and pressure accuracy is either +/- 0.01% FS or 0.05% FS depending on transducer used. | ||
Date | 2020-05-11T15:24:4 | ||
Identifier | SDN:L22::TOOL1437 | ||
Note | accepted | ||
Has Current Version | 1 | ||
version | 1 | ||
inScheme | http://vocab.nerc.ac.uk/scheme/SDNDEV/current/ | ||
Broader | L19:SDNKG17 |
sensor_model | Mapping: 1697092 |
L05 SeaDataNet device categories - (4) | |||
Related | L35:MAN0014 |
Valeport | Mapping: 1602479 |
B75:ORG00430 |
Valeport Ltd | Mapping: 1602473 |