Apollo SciTech AS-P3 combined pCO2 analyser and underway system
URI | http://vocab.nerc.ac.uk/collection/L22/current/TOOL1674/1/ | ||
---|---|---|---|
Within Vocab | SeaVoX Device Catalogue | ||
Alternative Labels | Apollo SciTech AS-P3 | ||
Definition | A continuous measurement pCO2 system. It is designed for land or ship-board laboratory use. Water sample is forced into the equilibrator by a shipboard pump. Equilibrated air is drawn from the equilibrator and dried with an electronic cooling system and a Nafion dry tube. It is then quantified with a CRDS CO2 analyser (Picarro G-2301) and/or an infrared CO2 analyser (LI-7000) before returning to the equilibrator. Optional sensors can be integrated. It measures at <100 -3000 uatm. Response time is 2 minutes. Repeatability is <1 ppm. Accuracy is +/- 1 to 2 ppm for pCO2. The detector can be replaced with a LI-850 analyser (pCO2 ranges 0-20,000 uatm) or LI-7815. | ||
Date | 2021-03-30T14:15:43 | ||
Identifier | SDN:L22::TOOL1674 | ||
Note | accepted | ||
Is Version Of | TOOL1674 | ||
version | 1 | ||
Broader | L05:351 |
dissolved gas sensors | |
L19:SDNKG17 |
sensor_model | ||
Related | L35:MAN0174 |
Apollo Scitech | |
B75:ORG01043 |
Apollo SciTech, LLC |